Tektronix THS730Aの取扱説明書
Service ManualTHS710A, THS720A,THS730A & THS720PTekScope070-9752-01This document applies to to serial numberB010100 and above and firmware version 1.00and above.First printing: February 1998WarningThe servicing instructions are for use by qualifiedpersonnel only. To avoid personal injury, do notperform any servicing unless you are qualified todo so. Refer to all safety summaries prior toperforming service.
ページ2に含まれる内容の要旨Copyright Tektronix, Inc. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes that in all previously published material. Specifications and price change privileges reserved. Printed in the U.S.A. Tektronix, Inc., P.O. Box 1000, Wilsonville, OR 97070–1000 TEKTRONIX and TEK are registered trademarks of Tektronix, Inc. Tek Secure is a registered trademark of Tektronix, Inc. TekTools, TekScope, and Isolate
ページ3に含まれる内容の要旨WARRANTY Tektronix warrants that this product will be free from defects in materials and workmanship for a period of three (3) years from the date of purchase from an authorized Tektronix distributor. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product. Batteries are excluded from this warranty. In order to obtain
ページ4に含まれる内容の要旨Service Assurance If you have not already purchased Service Assurance for this product, you may do so at any time during the product’s warranty period. Service Assurance provides Repair Protection and Calibration Services to meet your needs. Repair Protection extends priority repair services beyond the product’s warranty period; you may purchase up to three years of Repair Protection. Calibration Services provide annual calibration of your product, standards compliance and required audit documen
ページ5に含まれる内容の要旨Table of Contents General Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v Service Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ix Related Manuals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ix Conventions . . . . . . . . . . . . . . . . . . . .
ページ6に含まれる内容の要旨Table of Contents Options Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–1 Electrical Parts List Electrical Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8–1 Diagrams Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9–1 Mechanical Parts List Parts Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ページ7に含まれる内容の要旨Table of Contents List of Figures Figure 2–1: Removing and replacing the battery . . . . . . . . . . . . . . . . . 2–2 Figure 2–2: AC adapter attachment . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–2 Figure 2–3: Typical scope-mode display . . . . . . . . . . . . . . . . . . . . . . . . 2–3 Figure 2–4: Typical meter-mode display . . . . . . . . . . . . . . . . . . . . . . . . 2–4 Figure 2–5: Connector locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–8 Figure 3–1:
ページ8に含まれる内容の要旨Table of Contents List of Tables Table 1–1: Oscilloscope specifications . . . . . . . . . . . . . . . . . . . . . . . . . 1–1 Table 1–2: DMM specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–8 Table 1–3: General specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–10 Table 5–1: Summary of oscilloscope adjustment steps . . . . . . . . . . . . 5–5 Table 5–2: Summary of meter adjustment steps . . . . . . . . . . . . . . . . . 5–8 Table 6–1: Interna
ページ9に含まれる内容の要旨General Safety Summary Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use this product only as specified. Only qualified personnel should perform service procedures. To Avoid Fire or Connect and Disconnect Properly. Do not connect or disconnect probes or test leads while they are connected to a voltage source. Personal Injury Observe All Terminal Ratings. To avoid fire or shock hazard, obser
ページ10に含まれる内容の要旨General Safety Summary Symbols and Terms Terms in this Manual. These terms may appear in this manual: WARNING. Warning statements identify conditions or practices that could result in injury or loss of life. CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property. Terms on the Product. These terms may appear on the product: DANGER indicates an injury hazard immediately accessible as you read the marking. WARNING indicates an inju
ページ11に含まれる内容の要旨Service Safety Summary Only qualified personnel should perform service procedures. Read this Service Safety Summary and the General Safety Summary before performing any service procedures. Do Not Service Alone. Do not perform internal service or adjustments of this product unless another person capable of rendering first aid and resuscitation is present. Use Care When Servicing With Power On. Dangerous voltages or currents may exist in this product. Disconnect power, remove battery (if applicabl
ページ12に含まれる内容の要旨 ページ13に含まれる内容の要旨Preface This is the Service Manual for the THS710A, THS720A, THS730A, and THS720P TekScope instruments. This manual provides information to trouble- shoot and repair the instrument to the module level. Related Manuals Additional documentation for the instrument is contained in the related manuals listed below. User Manual Reference Programmer Manual Language Part Number Part Number Part Number English 070-9731-XX 070-9741-XX 070-9751-XX French 070-9732-XX 070-9742-XX German 070-9733-XX 070-9743-
ページ14に含まれる内容の要旨Preface Conventions In the Performance Verification, Adjustments, and Maintenance chapters of this manual, TekScope instrument setups are shown in tables similar to the one shown below. The header of each table contains icons that represent the controls and menu items used to set up the instrument. To make a specific setup, read the table from left to right and then from top to bottom as shown below. The table contains the symbol “—” if no action is required. 1. Choose scope 2. Press this but- 3
ページ15に含まれる内容の要旨 ページ16に含まれる内容の要旨Specifications This chapter contains the oscilloscope, DMM, and general specifications for the THS710A, THS720A, THS730A, and THS720P TekScope instruments. All specifications are guaranteed unless noted “typical.” Typical specifications are provided for your convenience but are not guaranteed. Specifications that are marked with the symbol are checked in the chapter Performance Verification. All specifications apply to all four instruments unless noted otherwise. All specifications assume hori
ページ17に含まれる内容の要旨Specifications Table 1–1: Oscilloscope specifications (Cont.) Inputs Maximum Voltage 600 V (CAT II) or 300 V (CAT III), using rated connectors or RMS RMS Between Common accessories and Earth Ground 30 V , 42.4 V , using noninsulated connectors and acessories RMS pk at BNC Maximum Voltage, 30 V , 42.4 V , using noninsulated connectors and acessories RMS pk Channel-to-Channel Commons Channel-to-Channel 100:1 at frequencies ≤50 MHz, measured on MATH Ch1 – Ch2 Common Mode waveform, with test signal
ページ18に含まれる内容の要旨Specifications Table 1–1: Oscilloscope specifications (Cont.) Vertical DC Gain Accuracy ±2% for Sample or Average acquisition mode Position Accuracy ±[0.4% × |(position × volts/div)| + (0.1 div × volts/div)] DC Measurement Measurement Type Accuracy A Accur racy y, A Aver erage e Absolute voltage measurements ±[2% × |reading + Acquisition Mode (position × volts/div)| + Using ≥16 Wave- (0.1 div × volts/div)] f forms rms Delta voltage between any two ±[2% × |reading| + waveforms acquired under
ページ19に含まれる内容の要旨Specifications Table 1–1: Oscilloscope specifications (Cont.) Internal Trigger Trigger Sensitivity, Coupling Sensitivity Ed dge e Tr Trigger er T Type ype, NOISE REJ 3.5 times the DC-coupled limits typ ypical HF REJ 1.5 times the DC-coupled limit from DC to 30 kHz, attenuates signals above 30 kHz LF REJ 1.5 times the DC-coupled limits for frequencies above 1 kHz, attenuates signals below 1 kHz Trigger Level Range ±4 divisions from center of screen Motor Trigger Level 0.1 to 5 divisions from cent
ページ20に含まれる内容の要旨Specifications Table 1–1: Oscilloscope specifications (Cont.) Measurements Cursors Voltage difference between cursors (V) Time difference between cursors (T) Reciprocal of T in Hertz (1/T) Phase difference between cursors (Degrees) Automated Amplitude, Burst Width, Cycle Mean, Cycle RMS, Fall Time, Measurements Frequency, High, Low, Max, Mean, Min, Negative Duty Cycle, Negative Overshoot, Negative Width, Pk – Pk, Period, Positive Duty Cycle, Positive Overshoot, Positive Width, Rise Time, an
# 取扱説明書 カテゴリ ダウンロード 1 Tektronix THS730A 取扱説明書 テレビゲーム 1 2 Tektronix THS720A 取扱説明書 工業用照明 64 3 Tektronix THS710A 取扱説明書 工業用照明 17 4 Tektronix VITS100 取扱説明書 工業用照明 1 5 Tektronix THS720P 取扱説明書 工業用照明 45 6 Edelbrock 25-9216 取扱説明書 工業用照明 0 7 3Com Switch 4500 26-Port 取扱説明書 工業用照明 69 8 3Com Switch 4500 PWR 50-Port 取扱説明書 工業用照明 7 9 3Com Switch 4500 PWR 26-Port 取扱説明書 工業用照明 17 10 Sony XDP-MU110 取扱説明書 工業用照明 1 11 3Com Switch 4500 50-Port 取扱説明書 工業用照明 38 12 Acclaim Lighting RGBX-DRUM-HIP 取扱説明書 工業用照明 2 13 Agilent Technologies 8164A 取扱説明書 工業用照明 6 14 Acclaim Lighting X-DRUM-HIP 取扱説明書 工業用照明 1 15 Agilent Technologies 85036B 取扱説明書 工業用照明 0 連絡先- contact -@- manualsbase.com